<3. Sensor Management>

3-1

IM 12A01S02-01E

1st Edition : May 10, 2013-00

3.

Sensor Management

Clicking on [Sensor management] from the toolbar will display “Current measurement readings,”

which shows the measurement value of sensors connected to the PC.
With this function, users can view sensor measurements, adjust sensor performance and

settings, or calibrate sensors.

By default, measurement targets and calibration objects are as shown in Table 3.1.

Table 3.1

Measurement targets and calibration objects

Sensor type *1 pH+ORP measurement *2 Measurement Target Calibration object *3

pH

Functions as a pH sensor

regardless of the setting

pH, Temperature

pH, Temperature

ORP *4

Functions as an ORP sensor

regardless of the setting

ORP, Temperature

ORP, Temperature

pH+ORP

pH

pH, Temperature

pH, Temperature

ORP *4

ORP, Temperature

ORP, Temperature

pH+ORP

pH, ORP, Temperature pH, ORP, Temperature

pH+rH

pH, rH, Temperature

pH, rH, Temperature

rH

rH, Temperature

rH, Temperature

*1:

“Sensor type” options of [Sensor management] – [Sensor setup].

*2:

“pH+ORP measurement” options of [Setup] – [pH/ORP] – [pH/ORP common settings].

*3:

“Calibration object” options of [Sensor management] – [Start calibration] – [Calibration wizard].

*4:

In the case of the ORP measurement on the basis of a reference electrode, please choose “ORP” of Sensor type “pH+ORP.”

In the case of the ORP measurement on the basis of a glass electrode, please choose Sensor type “ORP.”

For details, refer to User’s Manual of Model FLXA21 2-Wire Analyzer (IM 12A01A02-01E).

3.1 [Performance]

Clicking on [Performance] in “Current measurement readings” will display the sensor

performance screen, which shows the sensor information.

Table 3.2

Sensor performance

Item

Description

Calibration /

Maintenande

information

Last calibration pH, ORP, rH,

Temperature

Date/time of last calibration and type of calibration

Projectes replacement

Date of sensor replacement required for measurement accuracy,

which is predicted based on the preventative maintenance

function.

Calibration

data

pH

zero, slope

Calibration value and date/time of last update

ORP

zero, slope

rH

zero, slope

Temperature

offset

Sensor

wellness

Zero, Slope, Impedance 1,

Impedance 2

Sensor health

Higher % value of each gauge means superior parameter health.

Measuring

data

pH sensor

Electromotive force of the pH sensor

ORP sensor

Electromotive force of the ORP sensor

rH sensor

Electromotive force of the rH sensor

Impedance 1

Glass membrane resistance (for detecting possible damage to

the glass electrode)

Impedance 2

Impedance of the reference electrode junction