Yokogawa SPS24 User Manual

Page 31

Advertising
background image

<6. Setup>

6-5

IM 12A01S02-01E

1st Edition : May 10, 2013-00

Table 6.3

pH/ORP common settings (numbers in parentheses indicate related sections of FLXA21

User’s Manual (IM 12A01A02-01E))

Setting Item

Available Option

Initial Value

Remarks

pH+ORP measurement

pH, ORP, pH+ORP,

pH+rH, rH

pH+ORP

Time Axis Range(Max)

1 to 10 min.

5 min.

Timeline displayed

on the calibration

screen.

pH

calibration

settings

Zero and

Slope units

Zero unit

mV, pH

mV

Slope unit

%, mV/pH

%

Select buffer set

NIST/DIN 19266,

DIN 19267, US,

Programmable

NIST/DIN 19266

Setup

0.00 to 20.00 pH

(Table 6.4)

Appears when

selecting

“Programmable”.

Temperature difference

1.0 to 5.0 degC

-1.8 to 9.0 degF

2.0 degC

3.6 degF

Sets the maximum

temperature offset

between sensors for

multiple calibration.

Limits

Zero high

0.00 to 500.0 mV

0.00 to 8.45 pH

120.0 mV

2.03 pH

Zero low

-500.0 to 0.00 mV

-8.45 to 0.00 pH

-120.0 mV

2.03 pH

Slope high

100.0 to 110.0 %

59.16 to 65.08 mV/pH

110.0 %

65.08 mV/pH

Slope low

70.0 to 100.0 %

41.41 to 59.16 mV/pH

70.00 %

41.41 mV/pH

Step Range

0.01 to 1.00 pH

0.01 pH

Stabilization

time

2 to 30 s

10 s

ORP

calibration

settings

Limits

Zero high

0.00 to 500.0 mV

120.0 mV

Zero low

-500.0 to 0.00 mV

-120.0 mV

Slope high

100.0 to 110.0 %

110.0 %

Slope low

70.0 to 100.0 %

70.00 %

Step Range

1 to 100 mV

1 mV

Stabilization

time

2 to 30 s

10 s

rH

calibration

settings

Limits

Zero high

0.00 to 500.0 mV

120.0 mV

Zero low

-500.0 to 0.00 mV

-120.0 mV

Slope high

100.0 to 110.0 %

110.0 %

Slope low

70.0 to 100.0 %

70.00 %

Step Range

0.01 to 10.00 rH

0.05 rH

Stabilization time

2 to 30 s

10 s

Impedance

settings

Impedance 1 high limit

1.000 kOhm to

1.000 MOhm

(> Impedance 1 low

limit)

200.0 kOhm

When Impedance 1

is “low*.”

In the case of

“high,” use the

value of FINE limit

for impedance

1 in SENCOM

diagnostic criteria.

Impedance 1 low limit

1.000 kOhm to

1.000 MOhm

(< Impedance 1 high

limit)

1.000 kOhm

Impedance 2 high limit

1.000 kOhm to

1.000 MOhm

(> Impedance 2 low

limit)

200.0 kOhm

When Impedance 2

is “low*.”

In the case of

“high,” use the

value of FINE limit

for impedance

2 in SENCOM

diagnostic criteria.

Impedance 2 low limit

1.000 kOhm to

1.000 MOhm

(< Impedance 2 high

limit)

1.000 kOhm