2 [sensor setup, Sensor setup] -2, L[import setting – Yokogawa SPS24 User Manual

Page 19: L[export setting, L[initialize sensor

Advertising
background image

<3. Sensor Management>

3-2

IM 12A01S02-01E

1st Edition : May 10, 2013-00

3.2

[Sensor setup]

Clicking on [Sensor setup] in “Current measurement readings” will display the sensor setting

screen, where settings can be applied to each sensor individually.
For the settings common to pH sensors, go to the pH/ORP common setting in [Setup].

This screen displays values stored in connected sensors. Changed values will be displayed in

red. Click on [Synchronize] to send the new setting to SENCOM sensors.

l

[Import setting]

Imports sensor setting data (excluding Calibration) from a file.

l

[Export setting]

Exports sensor setting data (excluding Calibration) to a file.
The default file name is in the format of “Smart_SensorSetup(yymmddhhmmss).PH” which can

be replaced with up to 255 single-byte characters (127 double-byte characters). A file name

cannot contain any of the following characters: + * / ? ” < > |

l

[Initialize sensor]

Initializes the setting stored in SENCOM sensors.

Table 3.3

Sensor setting

Item

Range

Default

Remarks

Sensor

settings

Sensor type

pH, ORP, pH+ORP

pH+ORP

Calibration pH settings

Zero

-120.0 to 120.0 mV

4.97 to 9.03 pH

0.000 mV

7.00 pH

Unit can be selected at the

pH/ORP common settings.

Slope

70.00 to 110.0 %

41.41 to 65.08 mV/pH

100.0 %

59.16 mV/pH

Unit can be selected at the

pH/ORP common settings.

ITP

0.00 to 14.00 pH

7.00 pH

pH Auto correct

Disable, Enable

Disable

ORP/rH

settings

Zero

-120.0 to 120.0 mV

0.00 mV

Slope

70.00 to 110.0 %

100.0 %

Temperatutre

compensation

Compensation

Manual, Automatic

Automatic

Manual

temperature

0.0 to 100.0 degC

32.0 to 212.0 degF

25.0 degC

77.0 degF

Process temperature used

for manual compensation

Sensor

diagnostic

settings

Define

heat cycle

diagnostic

reference

Heat cycle

temperature

1 to 250 degC

34 to 482 degF

50 degC

124 degF

Heat cycle time

0.1 to 60.0 min.

10.0 min.

Define

SENCOM

diagnostic

reference

High pH value

-2.00 to 16.00 pH

13.00 pH

Low pH value

-2.00 to 16.00 pH

1.00 pH

Sterilization

temperature

0.0 to 155.0 degC

32.0 to 311.0 degF

155.0 degC

311.0 degF

Sterilization time

0.0 to 100.0 min.

100.0 min.

High temperature 1 -35.0 to 155.0 degC

-31.0 to 311.0 degF

155.0 degC

311.0 degF

High temperature 2 -35.0 to 155.0 degC

-31.0 to 311.0 degF

155.0 degC

311.0 degF