Lsetting for user defined buffer – Yokogawa SPS24 User Manual

Page 32

Advertising
background image

<6. Setup>

6-6

IM 12A01S02-01E

1st Edition : May 10, 2013-00

Temperature

compensation

Reference temperature

-30.0 to 140.0 degC

-22.0 to 284.0 degF

25.0 degC

77.0 degF

Process

Temp.

compensation

pH

None, Temp.

coefficient, Matrix,

NEN6411

None

Setup

-0.100 to 0.100 pH/

degC

-0.1 to 0.1 pH/degF

0.000 pH/degC

Appears when

“Temp. coefficient”

is selected.

Matrix settings.. Temperature range:

-30.0 to 140.0 degC

-22.0 to 284.0 degF

pH:

-2.00 to 20.00 pH

(Table6.5)

Appears when

“Matrix” is selected.

ORP

None, Temp.

coefficient

None

Setup

-10.00 to 10.00 mV/

degC

-5.6 to 5.6 mV/degF

0.000 mV/degC

Appears when

“Temp. coefficient”

is selected.

Error

display

Impedance 1 too high

off, Warnning, Fault

off

Impedance 1 too low

off, Warnning, Fault

off

Impedance 2 too high

off, Warnning, Fault

off

Impedance 2 too low

off, Warnning, Fault

off

Define

SENCOM

diagonostic

reference

FINE limit for impedance 1

1 to 10 MOhm

10.0 MOhm

When Impedance 1

is “high*.”

In the case of

“low,” use the value

of Impedance 1

high/low limit in

Impedance settings.

FINE limit for impedance 2

1 to 10 MOhm

10.0 MOhm

When Impedance s

is “high*.”

In the case of

“low,” use the value

of Impedance 2

high/low limit in

Impedance settings.

*:

In the case of FU20F pH/ORP SENCOM sensors, Impedance 1 is high and Impedance 2 is low.

l

Setting for user defined buffer

When “Programmable” is selected in pH calibration settings–Select buffer set, the user-defined

buffer solution configuration screen appears. Configuration can set for 3 buffer solutions.
Before configuring a new table, click on [Clear table] to clear all preset values.
After entering, execute [Check values] to check there are no errors.