Measurement, 1 precautions, Measurement -1 – Yokogawa SC72 Personal Handheld Conductivity Meter User Manual

Page 22: 1 precautions -1

Advertising
background image

IM 12D03D02-01E

3-1

3. Measurement

3. Measurement

3.1 Precautions

(1) Be sure to check that the cell constant and the temperature coefficient are set

correctly.

(2) Check that the plastic cover (for general-purpose sensor) and outer electrode (for

high purity water measurement) are secure.

(3) Do not use the SC72 meter to measure liquids with temperatures over 80

؇C. (If the

sensor grip is immersed, liquid temperature shall be below 50

؇C.) Do not use the

meter to measure extremely corrosive liquids such as solutions of hydrofluoric
acid.

(4) Remove dirt and stains from the meter with soft tissue. If necessary, wipe the

meter case with neutral detergent.

(5) If any problem with the meter arises during measurement, refer to the

Troubleshooting section later in this manual to determine the cause.

(6) After finishing measurement, flush stains on the sensor and measured solutions

with water, and store the meter (refer to Chapter 5, “Handling of the SC72
Personal Conductivity Meter”).

When the meter is used on a table:

The meter is designed as a portable instrument; however, to use it on a table, attach non-
slip pads (supplied with the instrument) at top and bottom of the meter to stop it from
moving when the sensor is moved.

SC METER

PERSONAL

2004.

SC72

S1.0

E00000 1

M O D EL

N o.

STYLE

Non-slip pads

F0301.EPS

Figure 3.1

Position of Non-slip Seats