2 washing the electrode, 2 washing the electrode -2 – Yokogawa SC72 Personal Handheld Conductivity Meter User Manual

Page 45

Advertising
background image

IM 12D03D02-01E

5-2

5. Handling of the SC72 Personal Conductivity Meter

5.2 Washing the Electrode

Dirt or stains on the electrode may have an adverse effect on the cell constant, thereby
making accurate conductivity measurement impossible. Therefore, after measurement,
rinse the electrode in clean water (for example, tap water) to remove stains. Even if
staining is not apparent, the sensor performance may have changed. If so, wash the
electrode by moving the sensor up and down in hydrochloric acid (about 0.1 mol/l) or
water with a little neutral detergent dissolved in it. [For general-purpose and pure-water
sensors: if the electrodes are difficult to clean, wipe them gently from top to bottom with
a cotton wool swab.] After cleaning the electrode, rinse it in water.

• General-purpose sensor

(The figure below shows the general-purpose sensor with the cover removed).

Remove the cover, and wipe stains off the electrodes (inner electrode and tip element)
with a cotton swab. Rinse in a cleaning solution consisting of water containing a little
neutral detergent, to remove stains from the electrode. After cleaning, wipe the electrode
element (inner and edge electrodes) with tissue paper or cotton wool.

Inner electrode (titanium)

Edge electrode (titanium)

F050201.EPS

• Sensor for high-purity water measurement

Remove the reference electrode part, and wipe the electrode elements (the shaded parts
in the figure) with tissue paper or cotton wool.

Reference electrode contact

Internal electrode

Reference electrode

Cotton wool swab

F050202.EPS

* Corrosion-resistant sensors and sensors for high-conductivity measurements

Use a beaker or the like containing water, dilute hydrochloric acid, or weak neutral
detergent; move the sensor up and down so it is immersed up to the air vent, and wash
in tap water to clean out the air vent.

Advertising