6. Troubleshooting

6. Troubleshooting

6.1 Causes of Abnormal Conductivity Display

The measured temperature is stable, but the conductivity reading is unstable, or an
unrealistic value is displayed, check the following:

(1) Is maintenance required, and has handling of the meter been adequate?

(2) Does the battery need replacement?

(3) Is SC72 meter unserviceable?

If any problem is evident, follow the procedure shown in Section 6.2 to determine the
cause and fix the problem. If you cannot fix it, contact your nearest Yokogawa sales
office.

Abnormal

Components need replacement

Displayed value unstable

Displayed value unrealistic

Error message displayed

Battery

Needs maintenance

Needs recalibration with standard solution

Electrode stained

Electrical leakage in connector

SC meter failure

Contact Yokogawa.

F0601.EPS

Figure 6.1

Causes of abnormality