2 detector – Yokogawa DM8C/VD6 Liquid Density Analyzer User Manual

Page 12

Advertising
background image

< 1. GENERAL >

1-2

IM 12T03A01-02E

1.1.2 Detector

(1) General Purpose Detector Model VD6D

Detector construction: Non-explosion protection, rain-proof construction
Case material:

Cast Aluminum alloy

Case coating:

Epoxy resin, baked fi nish

Case color:

Jade green (equivalent to Munsell 7.5BG4/1.5)

Wetted part materials:
Base:

SUS316

Vibrator:

SUS316 or Ni (Au Brazing: BAu·4)

Measuring liquid temperature:

–10 to 100 °C

Measuring liquid pressure:

2 MPa G or less

Maximum pressure:

4.9 MPa G

Steam tracing:

Available

Process connection:

Rc1/4

Electrical connection:

G3/4

Mounting:

2-inch. pipe mounting

Ambient temperature: –10 to 50 °C
Weight:

Approx. 12 kg

(2) Flameproof (Explosionproof) Detector Model VD6DF

Detector construction: TIIS; d2G3 or FM; Class I, Division 1, Groups C and D, Flameproof

construction

Process connection:

Rc1/4 or 1/4NPT female (only for VD6DF-*B/FM)

Electrical connection:

G3/4 or 3/4NPSM female (only for VD6DF-*B/FM)

Specifi cations are the same as for the (1) General Purpose Detector except for the above
construction.

(3) Sanitary Use Detector Model VD6DS

Process connection:

Special joint for connection to JIS 6A (6 mm dia.) pipe (within gasket)

Wetted part materials: Added to the standard model
Gasket:

Tefl on

O-Ring:

Viton

Stream tracing:

Not available

Specifi cations are the same as for the (1) General Purpose Detector except for the above two items.
Temperature detector protecting tubes are detachable.

Note: These detectors cannot be used with highly corrosive liquids and solutions likely to stick

to sensors. If it is desired to be applied to solutions containing slurry or sludge, consult with
YOKOGAWA. For measuring NaOH solutions, use sensors with a nickel vibrator.