< 3. INSTALLATION, PIPING AND WIRING >

3-13

IM 12T03A01-02E

3.4 Wiring

The vibration type density meter requires the following wirings:

(1) Wiring between analog output signal

(2) Wiring for analog output signal

(3) Wiring for digital output signal (10 m or less length is recommended).

(4) Wiring for contact output for abnormal status

(5) Wiring for power supplying

(6) Wiring for grounding

Figure 3.19 is a diagram of these wirings.

A2

A3

A4

A6

A5

B4

B5

A2

A3

A4

A6

A5

B4

B5

RED

BLUE

YELLOW

BROWN

GREEN

WHITE

GRAY

N.C

C

SG

RD

TD

+

-

+

-

G

L1 L2

Detector

Special cable

Converter

Power supply

Alarm

0 to 1 V DC

4 to 20 mA DC

RS-232-C signal

Closed when a
failure is detected or
the power fails.

‡Density signal after being

converted to the reference
temperature state

‡Density signal before being

converted to the reference
temperature state

‡7HPSHUDWXUHVLJQDO

Density signal after being
converted to the reference
temperature state

Figure 3.19

Wiring Diagram

The cable inlet of the density detector is located at the bottom of the terminal box. The density
converter has fi ve cable inlets and their size is 27 mm. A cable can pass any inlet of them (one cable
to one inlet). Figure 3.20 shows an example of allocated inlets.

Power supply

Contact output for abnormal status

Digital output signal

Inlet of special cable

Analog output signal

Figure 3.20

Cable Inlet of Density Converter