1 wiring between detector and converter, Wiring between detector and converter -14 – Yokogawa DM8C/VD6 Liquid Density Analyzer User Manual

Page 40

Advertising
background image

< 3. INSTALLATION, PIPING AND WIRING >

3-14

IM 12T03A01-02E

Figure 3.21 shows terminal arrangement of the density detector.

A2

A3

A5

B4

A4

A6

B5

Terminals

Function

B4

B5

A2

A3

A4

A5

A6

Output Frequency

Temperature Sensor

Power Supply for Pre-amp

Impressed Voltage for Pick-up Electrode

Earth Terminal

Figure 3.21

Terminal Arrangement of Density Detector

3.4.1 Wiring between Detector and Converter

This wiring is to be carried out with special cable of specifi ed length. The cable is terminated as
illustrated in Figure 3.22. Pay attention not to stain or wet terminated.

GRAY

YELLOW

BROWN

RED

RED

WHITE

WHITE

BLUE

BLUE

GREEN

GREEN

BROWN

YELLOW

GRAY

Density detector

Density converter

B4

B5

A4

A2

A3

A5

A6

B4

B5

A2

A3

A4

A5

A6

Figure 3.22

DM8W Cable

(Connection to Density Detector)

Remove the cover of the terminal box, using the spanner in accessories, and insert the special cable
into the cable inlet. Connect each conductor to the respective terminals.

A fl exible fi tting is used at the inlet of the density detector (note) and the cable from the detector runs
though metal conduit to the duct. For Model VD6DF explosionproof detector, wiring should be made
complying with the Recommended Practice.

The explosionproof fl exible fi tting for wiring should be used at the cable inlet (G3/4 female) of the
detector.

(Note) The fl exible fi tting is not inevitable. The Detector not moved for usual check or maintenance

does not require the fl exible tube.