< 1. GENERAL >

1-4

IM 12T03A01-02E

• Output

format

Measured

data

space

space

# * . * * * * _ * . * * * * _ * * * . *

C

R

L

F

(Note) Data are output in the order of density (data converted into that at a reference

temperature), density (at measuring temperature), and measuring liquid temperature.
Hold data (data immediately before entering the mode) is output during the time of
calibration or maintenance mode. However, holding is released immediately after
calibration or maintenance mode.

Calibration

status

space

# CALIBRATION _ START

C

R

L

F

(Note) Nothing is output during calibration. However, when a parameter error occurs, output

becomes as follows.

# ERROR - *

C

R

L

F

(*: 5 or 6)

space

# CALIBRATION _ END

C

R

L

F

Fail

alarm

space

# ERROR - * _ ERROR - *

C

R

L

F

(Note) When multiple errors occur, each error No. for the respective errors is output.

space

Example: # ERROR - 1 _ ERROR - 3

C

R

L

F

Output signal span:

0.05 to 0.5 g/cm

3

settable

Reference temperature set range:

0 to 100 °C (in increments or decrements of 1 °C)

Contact output on failure: One point. Contact closed on failure or power failure. Contact open

when

normal.

Permissible voltage:

220 V DC, 250 V AC

Permissible current:

2A (resistive load)

Permissible contact power:

60 W

Fault detecting contents: Detector failure and converter failure
Failure output:

Analog signal: Falls down to about –10 % of the output signal span

Digital signal:

Error message outputs

Output signal hold:

Holds in the CAL. or Maintenance mode.

Settable range for temperature coeffi cient:

0 to 0.002 g/cm

3

/°C

Calibration procedure: One-touch calibration by strong calibration liquid density

(onepoint

calibration)

Ambient temperature: –10 to 55 °C
Power supply:

90 to 132 V AC or 180 to 264 V AC, 50/60 Hz