2 density converter, 2 density, Converter – Yokogawa DM8C/VD6 Liquid Density Analyzer User Manual

Page 26: And temperature signal (voltage) v, Figure 2.3 block diagram of density converter

Advertising
background image

< 2. PRINCIPLES OF OPERATION >

2-4

IM 12T03A01-02E

2.2.2 Density

Converter

Figure 2.3 is a block diagram of the density converter.

As shown in the diagram, the density converter receives the density signal (frequency) F

ρ

and

temperature signal (voltage) V

T

from the density detector and fi nally outputs a 4 to 20 mA analog

signal and 0 to 1 V DC signal corresponding to the liquid density converted to that at reference
temperature. The digital signals of densities at measuring temperature and reference temperature,
and measuring temperature and measuring temperature value are also output. The converter circuit
comprises three printed circuit boards. Changing displayed items or setting constants are performed
by keys or transfer switches on the front panel.

CPU

Memory
ROM
RAM
EEPROM

Key &

Display

Controller

W.D.T.

PC

0 to 1 V

4 to 20 mA

Key

LED

TD

Digital Output

Fail Alarm

200 V

Reference frequency

Ref. temp. Zero

Frequency of Density (Fx)

Temp. Signal
(RT)

R/V

Ref. temp. Span

Multiplexer

Multiplexer

V/F

Integrator

V/I

Zero

Span

Driver

PC

PC

PC

COUNTER

P.W.
Modulation

Serial
Communi-
cation
Controller

Figure 2.3

Block Diagram of Density Converter

Advertising