< 4. OPERATION >

4-12

IM 12T03A01-02E

4.5 Operation

4.5.1 Operation

Procedure

Start the density measuring system, and monitor operating conditions for a while to confi rm that there
is no abnormality. Adjustment for the density measuring system during operation is unnecessary,
however, daily check of liquid fl ow rate and pressure is preferable. When the steam tracing pipe is
installed in the sampling unit, do not heat the detector.

If an error indication is displayed, make a countermeasure immediately. Errors are listed in Table 4.2.

Table 4.2

Error Table

Display

Contents

Error Release and Remedy

E001

Abnormal detector density signal

When cause is removed

E002

Abnormal detector temperature sinal

When cause is removed

E003

Abnormal on converter

E004

Overrang of measuring temperature

E005

Abnormal fi rst calibration

Check and reset the data required for

calibration.

E006

Abnormal second calibration

E007

Illegal data of EEPROM

Check and reset the loaded constant

(Note) E001, 002, 003, 004 and 007 are cancelled temporally by pressing the FUNCTION No. key

in maintenance or calibration mode, however in the measuring mode, the error indication is
displayed again. These errors are detected only in the measuring mode.

E005 and 006 occur only in the calibration mode and are displayed for one second when
calibration computation fi nishes and are cleared. After that, the density at the reference
temperature by the previous calibrating data is displayed.

4.5.2 Check of Measured Density Value

It is recommended that measured density values are checked at intervals of several weeks to
determine the period for calibration and cleaning.

(1) Measure the density using a hydrometer to compare it with the density indication (at the

reference temperature) on the vibration type density meter. When sampling the liquid, check the
pressure and the temperature are not largely different from those of the normal state.

(2) Measure density correctly at the reference temperature which is same as the sample liquid

temperature.

(3) Compare the density measured at (2) with the displayed value when sampling the liquid. If the

difference of them exceeds the allowable error range, clean the vibrator.

For cleaning procedure, refer to Section 5.1.1.

4.5.3 Shutdown

When shutdown sample following steps and the vibration type density meter also stops, drain the
sample from the detector.