< 3. INSTALLATION, PIPING AND WIRING >

3-15

IM 12T03A01-02E

Figure 3.23 Wiring to Density Detector Terminals

Figure 3.24

Spanner for Open/

closing

Terminal

Box

Cover

Cable

Terminal box case

Lock nut

G3/4 screw (TIIS explosionproof)
3/4NPSM (FM explosionproof)

Explosionproof flaxible fitting

Be sure to screw in
at least 12 mm (5 threads)

Figure 3.25

Explosionproof Flexible Fitting

(Connection to Density Converter)

The density converter is installed in fi eld or indoors (control room). The connector for passing through
cable should be sealed for air purge. The connector with a ring packing as shown in Figure 3.26
should be used, or the inlet of metal conduit is fi lled with silicon sealing compound. Duct or protective
pipe for the cable should be provided to prevent the cable from damage.

Cable

Converter case

Lock nut

G3/4 screw

Flaxible fitting

Nut

Figure 3.26 Packing Gland

Type Connector for

Passing through Cable

Figure 3.27

Cable Inlet for Metal Conduit

Take the special cable into the converter from the leftmost inlet and connect the each conductor to the
respective terminals.