Chapter 4

Calibrator

4.1 Connecting the MW100 Data Acquisition Unit

Notes before Using the Calibrator

• If connected to the MW100 from a browser, drop the connection before connecting

with the Calibration Software. Also, place the MW100 in Calibration mode. To enter
Calibration mode, turn OFF the power to the MW100 main unit, then turn it back ON
again while holding down the

USER 1 key. Release the USER 1 key one second or

more after the self check begins (the 7-segment LED displays

). To exit

Calibration mode, turn the power OFF.

• If you change the module configuration before connecting with the Calibration

Software, carry out system reconfiguration on the browser (see section 3.3,
“Configuring the System” in the MW100 Data Acquisition Unit User’s Manual, IM
MW100-01E).

Procedure

1.

Start the Calibrator.

The Unit Information window opens.

2.

In the IP Address/Host Name box, enter the IP address or host name of the

MW100 to be calibrated.

3.

Click Connect.

Enter the IP address

or host name of the

MW100 to be calibrated.

Execute connection