4-11

IM MW180-01E

Calibrator

1

2

3

4

5

App

Index

• 6-Channel, Medium-Speed, Four-Wire RTD Resistance Input Module (MX110-V4R-M06)

Measurement Range to Be Calibrated

Input Value 1

Input Value 2

20 mV

0 mV on Ch1

20 mV on Ch2

60 mV

0 mV on Ch1

60 mV on Ch2

200 mV

0 mV on Ch1

200 mV on Ch2

1 V

0 V on Ch1

1 V on Ch2

2 V

0 V on Ch1

2 V on Ch2

6 V

0 V on Ch1

6 V on Ch2

20 V

0 V on Ch1

20 V on Ch2

100 V

0 V on Ch1

100 V on Ch2

RTD(1 mA)20 mV

0 Ω on Ch3

20 Ω on Ch4

RTD(1 mA)60 mV

0 Ω on Ch3

60 Ω on Ch5

RTD(1 mA)200 mV

0 Ω on Ch3

200 Ω on Ch6

RTD(1 mA)600 mV

0 Ω on Ch3

300 Ω on Ch4

RTD(0.25 mA)600 mV

0 Ω on Ch3

2400 Ω on Ch5

RTD(0.25 mA)1 V

0 Ω on Ch3

3000 Ω on Ch6

• 4-CH, Medium-Speed Strain Input Module (MX112-B12-M04, MX112-B35-M04,

MX112-NDI-M04)

Measurement Range to Be Calibrated

Connection Value 1

Connection Value 2

2000 µSTR (strain)

120.000 Ω on Ch2

117.154 Ω on Ch2

20000 µSTR (strain)

120.000 Ω on Ch2

113.010 Ω on Ch2

200000 µSTR (strain)

120.000 Ω on Ch2

80.000 Ω on Ch2

Use a 4-gauge method connection when performing calibration. For information on

this connection, see section 2.4 of the MW100 Data Acquisition Unit User’s Manual (IM
MW100-01E).

• 8-CH, Medium-Speed Analog Output Module (MX120-VAO-M08)
Adjust so that all channels output 0 V and 10 V. This differs from other input modules.

Range to Be Calibrated

Output Value 1

Output Value 2

10 V

0 V

10 V

Note

• For a description of the measurement range and accuracy of each input module, see

chapter 5 in the MW100 Data Acquisition Unit User’s Manual (IM MW100-01E).

• For procedures and wiring for calibration, see the MW100 Data Acquisition Unit User’s

Manual (IM MW100-01E), section 4.4.

4.2 Calibration Procedure