Explanation, Calibrated parameters – Yokogawa PC-Based MX100 User Manual

Page 59

Advertising
background image

4-10

IM MW180-01E

4.2 Calibration Procedure

Explanation

Calibrated Parameters

• 4-CH, High-Speed Universal Input Module (MX110-UNV-H04)

Measurement Range to Be Calibrated

Input Value 1

Input Value 2

20 mV

0 mV

20 mV

60 mV

0 mV

60 mV

200 mV

0 mV

200 mV

1 V

0 V

1 V

2 V

0 V

2 V

6 V

0 V

6 V

20 V

0 V

20 V

100 V

0 V

100 V

RTD(1 mA)200 mV

0 Ω

200 Ω

RTD(1 mA)600 mV

0 Ω

300 Ω

RTD(1 mA)1 V

0 Ω

500 Ω

RTD(2 mA)60 mV

0 Ω

30 Ω

RTD(2 mA)200 mV

0 Ω

100 Ω

RTD(2 mA)600 mV

0 Ω

300 Ω

RTD(2 mA)1 V

0 Ω

250 Ω

• 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10)

Measurement Range to Be Calibrated

Input Value 1

Input Value 2

20 mV

0 mV on Ch1

20 mV on Ch2

60 mV

0 mV on Ch1

60 mV on Ch2

200 mV

0 mV on Ch1

200 mV on Ch2

1 V

0 V on Ch1

1 V on Ch2

2 V

0 V on Ch1

2 V on Ch2

6 V

0 V on Ch1

6 V on Ch2

20 V

0 V on Ch1

20 V on Ch2

100 V

0 V on Ch1

100 V on Ch2

RTD(1 mA)60 mV

0 Ω on Ch3

60 Ω on Ch6

RTD(1 mA)200 mV

0 Ω on Ch3

200 Ω on Ch4

RTD(1 mA)600 mV

0 Ω on Ch3

300 Ω on Ch5

• 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30)

Measurement Range to Be Calibrated

Input Value 1

Input Value 2

20 mV

0 mV on Ch1

20 mV on Ch2

60 mV

0 mV on Ch1

60 mV on Ch2

200 mV

0 mV on Ch1

200 mV on Ch2

1 V

0 V on Ch1

1 V on Ch2

2 V

0 V on Ch1

2 V on Ch2

6 V

0 V on Ch1

6 V on Ch2

20 V

0 V on Ch1

20 V on Ch2

100 V

0 V on Ch1

100 V on Ch2