8. CLEANING

To clean the tray and housing of the scales use a
household detergent or standard disinfectant as
required. Take note of the instructions provided by the
manufacturer.

9. WHAT TO DO IF …?

Malfunction

Cause/Remedy

… no weight is dis-
played during weighing?

The scales are not connected to the power supply

- Check the scales are switched on
- Check batteries are inserted

0.000

does not appear

before weighing?

A load was placed on the scales before being
switched on.

- Remove load on scales
- Switch scales off and back on again

… a segment lights up
continuously or not at
all?

There is a fault at that point.

- Notify maintenance service.

is displayed?

Battery voltage is falling off.

- Insert new batteries as soon as possible

bAtt

is displayed?

Batteries are flat.

- Insert new batteries

St0p

is displayed?

The maximum load has been exceeded.

- Reduce load

is

displayed?

The ambient temperature of the scales is too high
or too low.

- Set up scales in an ambient temperature

between +10 °C and +40 °C

- Wait approx. 15 minutes until scales have

adjusted to ambient temperature