If you wish to check that the scales are properly calibrat-
ed, proceed as follows:
1. Switch the scales off if necessary.
2. Hold down any key and start up the scales.
The current content of the calibration counter will
flash in the display for a few seconds.
3. Compare the output content of the calibration
counter with the figure on the calibration counter
sticker.
The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, recali-
bration is necessary. Contact your service partner or the
seca customer service department. Should recalibration
be necessary, the seca calibration counter sticker
shown above will be replaced by the adjacent recali-
bration sticker to identify the calibration counter read-
ing. This sticker will be secured with an additional seal
by a person authorised to perform recalibration. The
recalibration sticker is available from the seca customer
service department under the number 14-05-01-886.
11. TECHNICAL DATA
Technical data for
seca 376
Dimensions
• Depth
• Width
• Height
358 mm
620 mm
190 mm
Weight
3.5 kg
Temperature range
+10° C to +40°C
Height of figures
21 mm
Power supply
Battery
Power supply unit (optional)
Power consumption
• with wireless module disabled
• with wireless module activated
approx. 32 mA
approx. 50 mA
Maximum runtime in battery mode
• with wireless module disabled
• with wireless module activated
approx. 5,600 minutes
Power supply unit recom-
mended