Seca 376 User Manual

Page 57

Advertising
background image

What to do if …? • 57

English

... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?

• The device was unable to send measured

results to the radio receiver (seca wireless prin-
ter or PC with seca USB wireless module).

– Check that the scales are integrated in

the wireless network.

– Check that the receiver is switched on.

• Nearby RF equipment (e.g. mobile phones) are

interfering with reception.
- Make sure that RF equipment is kept at least

1 metre away from transmitters and receivers
in the seca wireless network.

NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.

... only the item "SYS" is
visible in the rf menu?

• The wireless module is disabled.

- Activate wireless module (see "Activate

wireless module (SYS)" on page 52).

... only the items "SYS"
and "lrn" are visible in
the rf menu?

• The wireless module is activated but no wireless

group set up.
- Set up wireless group (see "Set up wireless

group (Lrn)" on page 52).

... the items "APrt" and
"Time" are not visible in
the rf menu?

• No wireless printer is logged into the wireless

group.
- Use menu item "lrn" to log the wireless printer

into the wireless group (see "Set up wireless
group (Lrn)" on page 52).

Er:X:11

is displayed?

The scales are too high, or too great a load is
applied at one point.

- Reduce load on scales or distribute weight

more evenly

- Restart scales

Er:X:12

is displayed?

The scales have been switched on with too great
a load.

- Reduce load on scales
- Restart scales

Er:X:16

is displayed?

Oscillation of the scales has occurred, preventing
determination of the zero point.

- Do not touch the trolley or table supporting

the scales

- Restart scales

Malfunction

Cause/Remedy

Advertising