Before you get started, 1 installation of column – Seca 704 User Manual

Page 46

Advertising
background image

46 •

5. BEFORE YOU GET STARTED …

5.1

Installation of column

The column with the LCD display can be fitted in two
positions, so allowing you to determine the direction in
which the display is legible - facing either towards or
away from the platform.

1. Fit the column cover to the column.

2. Place the column in the recess, facing it in the

desired direction.

3. Screw the column to the weighing platform from

underneath.
If necessary, lay the scales down carefully on their
side to tighten up the screws.

4. On the underside of the weighing platform connect the

plugs of the display cable as shown in the diagram.

ATTENTION!

Malfunction with any error in installation

If cables are strained during fitting, e.g. bent
tight or with plugs kinked, this may result in mal-
functions and failure of the display.
– Make sure that all cables are laid so as to

prevent them from being bent too tight or the
plugs kinked.

– Ensure strain relief by laying all cables in the

relevant holders.

5. Secure cables in their holders.

6. Then place the scales upright again.

7. Slide the column cover right down until it engages

on the weighing platform.

Advertising