10.MAINTENANCE/RECALIBRATION

10.1 Information about maintenance and recalibration

We recommend having maintenance performed prior
to recalibration of the scales.

ATTENTION!

Incorrect measurement due to improper
maintenance

– Make sure that maintenance and repair are

only carried out by an authorised service
partner.

You can find your local service partner at

www.seca.com or just send an e-mail to
[email protected].

Ensure that recalibration is carried out by authorised
persons in accordance with the national legal require-
ments. The year in which initial calibration was per-
formed can be found behind the CE mark on the rating
plate above the number of the notified body 0109
(Hesse Weights & Measures Office).

Recalibration is required in all cases if one or more cali-
bration seals have been damaged or the content of the
calibration counter does not match the number on the
valid calibration counter sticker.

Er:X:16

is displayed?

Oscillation of the scales has occurred, preventing
determination of the zero point.

- Restart scales

… when the Enter key
(send/print) is pressed
and

Er:X:71

is dis-

played?

Data transmission not possible, wireless module is
disabled.

- Activate wireless module (see "Activate

wireless module (SYS)" on page 58).

… when the Enter key
(send/print) is pressed
and

Er:X:72

is dis-

played?

Data transmission not possible, no wireless group
set up.

- Set up wireless group (see "Set up wireless

group (Lrn)" on page 59).

Malfunction

Cause/Remedy