Continuous display of measured result (hold) – Seca 704 User Manual

Page 49

Advertising
background image

Operation • 49

English

4. Hold down the arrow key (hold/2 in 1) until NET

appears in the display.
The weight is stored.

0.0

appears in the display.

ATTENTION!

Incorrect measurement due to change in initial
weight

If the child is weighed with a different initial
weight, the child's weight will not be calculated
correctly.
– Make sure that the child is always weighed

with the adult used to determine the initial
weight..

– Check that the weight of this adult does not

change, e.g. by taking off a garment.

5. Ask the adult to step onto the scales with the child.

The child's weight is displayed.
The

symbol, the

symbol and the messages

"HOLD" and "NET" are displayed.

6. Ask the adult to step off the scales with the child.

7. To disable the 2 in 1 function, press the arrow key

(hold/2 in 1) until the

symbol, the

symbol and

and the messages "HOLD" and "NET" are no longer
displayed or switch the scales off.

Continuous display of

measured

result (HOLD)

If you activate the HOLD function, the weighing value
will continue to be displayed after the load has been
removed. This allows you to attend to the patient
before noting down the weight.

1. Check that there is no load on the scales.

2. Ask the patient to step onto the scales.

3. Briefly press the arrow key (hold/2 in 1).

The display will flash until a stable weight is meas-
ured. The weighing value will then be displayed
continuously. The

symbol (non-calibratable func-

tion) and the message "HOLD" will be displayed.

4. To disable the HOLD function, briefly press the ar-

row key (hold/2 in 1).
The

symbol and the message "HOLD" will no

longer be displayed.

Advertising