Delmhorst Instrument TotalCheck User Manual

Page 13

Advertising
background image

Copyright

2006, Delmhorst Instrument Co.

Page 13 of 32

1. To change the temperature, use the



keys from the CREATE JOB sub menu

to highlight the temperature, T: 70ºF.

2. Press the SELECT key to enter the COMP TEMP sub menu.

3. Using the



keys, scroll up or down to a higher or lower temperature value.

Temperature values will change in 5º F increments.

4. To change between C and F, press the SELECT key when the COMP TEMP is

highlighted.

TO SET THE ALARM IN ENHANCED PIN MODE:


1. If you wish to set the alarm to beep after a pre-selected value is reached, use the

key from the MC READ screen, highlight ALARM, and press the SELECT key to

enter the ALARM sub menu.


2.

The default status of the alarm is “OFF.” Press the SELECT key to turn it on.

4. After the alarm is turned on, use the



keys to scroll up or down to a higher or

lower alarm value.

5. After you have selected the alarm value, use the



key to highlight DONE.

Press SELECT to get back to the MC READ screen


Default alarm settings in PIN mode are as follows for different materials:


All wood materials:

15%

Drywall:

1.0%

Concrete:

68 (on the 0-100 relative scale)


TO TAKE PIN READINGS IN ENHANCED MODE:

1. From the Enhanced Main Menu, use the



keys to highlight #1, PIN MODE.


2. Press the SELECT key to get into the ENHANCED READ screen. The current

location will stay the same while toggling between scan and pin modes.

Advertising