2. Ensure correct polarity, and push the battery in flush with the bottom board until the

connectors snap together on both sides.

3. Replace the battery compartment lid.

IMPORTANT NOTE:

DO NOT REMOVE THE BATTERY WITHOUT TURNING THE METER OFF FIRST.

BATTERY REMOVAL WHILE THE METER IS ON MAY CAUSE LOSS OF STORED

READINGS AND MAY CORRUPT THE STORAGE MEMORY. IF THIS OCCURS,

CERTAIN FUNCTIONS (SUCH AS STATISTICS, READING RECALL, ETC) WILL FAIL

AND LOCK UP THE METER. If the battery is removed by accident while the meter is

ON, and if the readings already stored are important, DO NOT store additional readings.

Turn the meter back on and perform a PC download. Beware that any readings stored

during the power cycle that included the battery removal WILL NOT be available and

CANNOT be recovered. Once the download is complete, perform a meter reset by

choosing “RESTORE DEFAULTS” in the Storage Menu. If a download is not necessary,

go directly to “RESTORE DEFAULTS”.

LOW BATTERY:

The meter features a battery status monitor, designed to warn the user as well as protect

measurement accuracy from impending battery failure conditions. The battery warning is

triggered by either continuous or temporary low voltage conditions. Visible (a battery

icon on the top right side of the display) and audible (buzzer warble) indicators

accompany a battery warning.

Once a permanent low battery condition is detected, all measurement functions are

disabled. Enough power remains for other menu functions, including a PC download,

although almost any user action at this time causes a low-battery warning to be

generated. The battery should be replaced immediately. If the battery reaches critical

levels, the meter will refuse to stay on at power-up.

This gradual warning system is intended to provide the user advanced battery status

notice and give ample time for replacement before operational limitations occur.

NOTE: Stored readings are not lost during battery replacement or low battery conditions.

Current meter settings are maintained as well.