View system information, View system error, E128) – GW Instek GSP-830 User Manual User Manual

Page 128: Amp uncal, Med unlock, Ext unlock, Lo1 unlock, Lo3 unlock, Ref unlock

Advertising
background image

GSP-830 User Manual

128

3. Press F2 (Self Test).

Self Test...

F 2

4. The GPIB result appears at F1.

If the result is Fail (underlined),
contact service personnel.

GPIB

Pass Fail

F 1

GPIB constraints

Keep these rules when using the GPIB interface.

•

Altogether less than 15 devices & 20m cable length,
2m between each device on the bus

•

Unique address assigned to each device

•

At least 2/3 of the GPIB devices turned on

•

No loop or parallel structure allowed

View System Information

View system error

Panel operation

View the bottom of the screen, the error message area.
If there is a system error, the message appears in red.

Center : 1.5GHz

(EXT Unlock)

EXT

Unlock

Amp

Uncal

•

Inappropriate RBW or VBW is selected.

•

Frequency is less than 15MHz and
amplitude is less than −30dBm.

EXT

Unlock

External reference input is not working
properly.

LO1

Unlock

Local oscillator 1 is not working properly.

LO3

Unlock

Local oscillator 3 is not working properly.

Med

Unlock

±

1ppm stability signal is not working

properly. Appears only when the optional
±

1ppm Stability Module is installed.

Ref

Unlock

Internal reference signal is not working
properly.