GSP-830 User Manual

22

Error Check

This section assumes that the GSP-830 is already powered up (page20).

1. Check system
error

Check for error messages at the bottom of the display,
next to the command window.

Center : 1.5GHz

(EXT Unlock)

EXT

Unlock

Contact the service center if any of the following
messages remains in the display.

Amp

Uncal

•

Inappropriate RBW or VBW is selected.

•

Frequency is less than 15MHz and
amplitude is less than −30dBm.

EXT

Unlock

External reference input is not working
properly.

LO1

Unlock

Local oscillator 1 is not working properly.

LO3

Unlock

Local oscillator 3 is not working properly.

Med

Unlock

±

1ppm stability signal is not working

properly. Appears only when the optional
±

1ppm stability module is installed.

Ref

Unlock

Internal reference signal is not working
properly.

2. Check self test
result

View the GSP-830’s self-diagnosis test result. Press the
System key→F6 (More) →F2 (Self Test).

Self Test...

F 2

More...

F 6

System

The test automatically runs at each power-up. The
underline shows the result, pass or fail. Contact the
service center if any of the items fails.

GPIB

Pass Fail

F 1

GPIB module connectivity
(available only when installed)

Flash

Pass Fail

F 2

Internal Flash memory for storing
the system code/data