Location: indoor, Relative humidity: < 85, Temperature: 0°c to 40°c – GW Instek GSP-830 User Manual User Manual

Page 7

Advertising
background image

SAFETY

INSTRUCTIONS

7

•

Use only mating connectors, not bare wires, for the

terminals.

•

Do not block or obstruct the cooling fan vent opening.

•

Do not perform measurement at a power generating source
or building installation site (Note below).

•

Do not disassemble the GSP-830 unless you are technically
qualified.

(Note) EN 61010-1:2001 specifies the measurement categories and
their requirements as follows. The GSP-830 falls under category II.

•

Measurement category IV is for measurement performed at the
source of low-voltage installation.

•

Measurement category III is for measurement performed in the
building installation.

•

Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.

Power Supply

WARNING

•

AC Input voltage: 100 to 240 V AC, 50/60Hz, 90W
maximum

•

DC Input voltage: 12V DC, 40W maximum

•

The power supply voltage should not fluctuate more than

10%.

•

Connect the protective grounding conductor of the AC

power cord to an earth ground, to avoid electrical shock.

Battery

CAUTION

•

Rating: 11.1V lithium-ion battery pack x 2

•

Turn off the main power switch before installing or taking

out the battery packs.

Fuse

WARNING

•

Fuse type: T1.6A/ 250V, time-delay, 5x20mm glass

•

To ensure fire protection, replace the fuse only with the

specified type and rating.

•

Disconnect the power cord before fuse replacement.

•

Make sure the cause of fuse blowout is fixed before fuse
replacement.

Cleaning the
GSP-830

•

Disconnect the power cord before cleaning.

•

Use a soft cloth dampened in a solution of mild detergent

and water. Do not spray any liquid into the GSP-830.

•

Do not use chemicals or cleaners containing harsh
products such as benzene, toluene, xylene, and acetone.

Operation
Environment

•

Location: indoor, no direct sunlight, dust free, almost
non-conductive pollution (Note below)

•

Relative Humidity: < 90%

•

Altitude: < 2000m

•

Temperature: 18°C to 28°C

Storage
Environment

•

Location: indoor

•

Relative humidity: < 85%

•

Temperature: 0°C to 40°C