PEL-2000 Series User Manual
162
Setting (OCP/OVP/OPP/UVP)
Background
Over Protection is used to set the voltage, current
or power limit. In the event that the current,
voltage or power exceeds the over protection
settings, the load module alarm will show an error
message and beep an alarm.
When tripped, Under Voltage Protection (UVP)
will turn off the load. UVP trips when the load
voltage drops below a set limit.
Only when the protection settings are set to On
(XXP Setting -On) will the protection modes be
active.
All protection settings can be set to 2% higher than
specification rating.
Parameters
OCP Level
0~Rating A +2%
OCP Setting
ON/OFF/Clear
OVP Level
0~RatingV+2%
OVP Setting
ON/OFF/Clear
OPP Level
0~RatingW+2%
OPP Setting
ON/OFF/Clear
UVP Level
0~RatingV+2%
UVP Setting
Clear
Panel operation
Ensure the menu is the configuration
menu. See page 160.