Configure go/nogo interface – GW Instek PEL-2000 Series User Manual User Manual
Page 239

INTERFACE
239
Connection
Input: active low (0-1V), active high (4-5V)
Inputs (5V, 10kΩ pull-up resistor)
Open collector outputs (Load status, Alarm
Status, +5V) maximum 30VDC with 1.1V
saturation voltage (100mA).
When enable is on (active low), the following is
disabled from the mainframe: activating loads
and recalling preset or setup memory.
Frame Link
constraints
Maximum 5 (1 master + 4 slave units) devices
can be linked altogether with a maximum cable
length of 30cm for each cable.
All the connected devices must be turned on.
No loop or parallel connections
Configure Go/NoGo Interface
Connection
Use a DSUB (DB-15 Female)
connector to connect to the
Go/NoGo port.
The Go/NoGo port is an
output only port.
Pin assignment
Pin1
Ch1_GO/NG
Pin9
Ch5_GO/NG
Pin2
GND
Pin10
GND
Pin3
Ch2_GO/NG
Pin11
Ch6_GO/NG
Pin4
GND
Pin12
GND