Load module
considerations
The PEL-2000 supports single and dual channel
load modules.
Single channel load modules have one bank of
negative terminals and one bank of positive
terminals. Each terminal pair has a 40A capacity.
For higher loads, each terminal can be wired in
parallel to increase capacity.
Dual channel load modules have one bank of
positive and negative terminals for each channel.
Single Channel Load
Module
Dual Channel Load
Module
L
80V DC
MAX MAX
80V DC
R
Left channel
Right channel
Connection
Follow the procedure below for all load
connections.
CAUTION
Ensure that power is off from the load generator and
the DUT before making any connections.
Steps
1. Carefully lift the terminal covers.
2. Connect the positive (+) terminal on the load
module to the high potential output of the
DUT.
3. Connect the negative (-) load terminal to the
low potential output of the DUT.