Specifications
General
Note
All specifications are ensured only under a single display.
At least 30 minutes of warm-up time is required before
applying these specifications.
Make sure the power ground is connected.
Resolution
Type
Digit
Slow (S)
5 ½ Digit
Medium (M)
4 ½ Digit
Fast (F)
3 ½ Digit
Operation
Environment
Ambient Temperature 0°C ~ 40°C, Relative Humidity < 75%
(For full accuracy: 18°C ~ 28°C)
Temperature
Coefficient
< 0.2 x applicable accuracy per degree (°C)
(for 0°C ~ 18°C and 28°C ~ 40°C)
Storage
Environment
Ambient Temperature −10°C ~ 70°C
Relative Humidity: 0°C ~ 35°C < 75%, 35°C ~ 50°C < 50%
Power Source
AC 100–240V ± 10%, 50–60Hz
Dimension
265(W) x 107(H) x 350(D) mm
Weight
Approx. 2.6kg without option
Reading rates (readings/sec)
Function
Rate
S
M
F
DCV
10
30
60
DCI
10
30
60
ACV
1
5
20
ACI
1
5
20
2/4WΩ
(10M/100MΩ)
1
1.5
2
2/4WΩ (others) 3
5
8
ACV+DCV
0.5
1
3
ACI+DCI
0.5
1
3
Diode
30
30
60