Dbm/db measurement, Measure dbm, E43) – GW Instek GDM-8200A User Manual User Manual
Page 43

ADVANCED MEASUREMENT
43
dBm/dB Measurement
Applicable to
(NOT applicable to ACV+DCV)
Background
Using the ACV or DCV measurement result, the
GDM-8200A series calculates the dB or dBm value based
on a reference resistance value in the following way.
dBm
10 x log
10
(1000 x Vreading
2
/ Rref)
dB
dBm – dBmref
Parameters
Vreading
Input Voltage, ACV or DCV
Vref
Reference voltage obtained by Rref/1mW
Rref
Reference resistance simulating an output
load
dBmref
Reference dBm value
Measure dBm
Activate dBm
Press the Shift key followed by the
key. The 1st display shows dBm,
and the 2nd display shows the
reference resistance.
dBm result
appears
dBm
Indicates dBm measurement
600Ω
2nd display indicates the reference resistance
Select
reference
resistance
To change the reference resistance,
press the Up/Down key. The new
resistance appears in the 2nd display.
The following is the resistance list.
2
4
8
16
50
75
93
110
124
125
135
150
250
300
500
600
800
900
1000
1200
8000