Set trigger delay – GW Instek GDM-8200A User Manual User Manual
Page 60

GDM-8200A Series User Manual
60
1. Activate
external trigger
Press the Shift key followed by the
TRIG key. The EXT indicator
appears on the display.
2. Start trigger
Press the TRIG key to start
triggering manually. The
indicator turns On.
Reading indicator
The reading indicator does not flash before triggering
(can be on or off). After triggering, the indicator flashes
according to the external signal trigger timing.
Exit external
trigger
Press the Shift key followed by the
TRIG key. The EXT indicator
disappears and the trigger goes back
to internal mode.
Set trigger delay
Background
Trigger delay defines the time rag between triggering and
measurement start. The default is set at 10ms.
Panel operation
1. Press the Shift key, the 2ND
(Menu) key, the Right key, the
Down key. The delay menu
appears.
2. Press the Down key. The delay
setting appears.