Diode/continuity, Frequency, Temperature – GW Instek GDM-8200A User Manual User Manual
Page 119

APPENDIX
119
100.000MΩ 120.000MΩ
199.999MΩ
3.0% + 8
M
100.00Ω
120.00Ω
199.99Ω
0.05% + 5
1.0000kΩ
1.2000kΩ
1.9999kΩ
0.05% + 3
10.000kΩ
12.000kΩ
19.999kΩ
0.05% + 3
100.00kΩ
120.00kΩ
199.99kΩ
0.05% + 3
1.0000MΩ 1.2000MΩ
1.9999MΩ
0.05% + 3
10.000MΩ 12.000MΩ
19.999MΩ
1.5% + 3
100.00MΩ 120.00MΩ
199.99MΩ
5.0% + 5
F
100.0Ω
120.0Ω
199.9Ω
0.05% + 2
1.000kΩ
1.200kΩ
1.999kΩ
0.05% + 2
10.00kΩ
12.00kΩ
19.99kΩ
0.05% + 2
100.0kΩ
120.0kΩ
199.9kΩ
0.05% + 2
1.000MΩ
1.200MΩ
1.999MΩ
0.05% + 2
10.00MΩ
12.00MΩ
19.99MΩ
1.5% + 2
100.0MΩ
120.0MΩ
199.9MΩ
5.0% + 2
Diode/Continuity
Note
Max. Input: 500V DC or 500V rms AC
Item
Range
Diode
Approx. 2V, 0.5mA
Continuity
1 ~ 1000Ω
Frequency
Note
Max. Input: 750V rms or 1000V peak
Frequency
Sensitivity
Accuracy (reading%+digits)
10Hz ~ 100kHz 0.1V
0.05% + 15
100kHz ~ 600kHz 1V
0.05% + 3
600kHz ~ 800kHz 2.5V
0.05% + 3
Temperature
Note
Sensor errors excluded from Temperature specifications
Type
Measurement Range
Thermo Couple
K
0 ~ +300°C
T
0 ~ +300°C
J
0 ~ +300°C
Resolution
0.01°C (0 ~ 300°C)