Relative value measurement, E46) – GW Instek GDM-8200A User Manual User Manual
Page 46

GDM-8200A Series User Manual
46
Relative Value Measurement
Applicable to
( )
+
( )
+
Background
Relative measurement stores a value, typically the data at the
moment, as the reference. The following measurement is
shown as the delta between the reference.
1. Activate
Relative
measurement
Press the REL key. The
measurement reading at the
moment becomes the
reference value.
2. Relative
measurement
display
appears
REL
Indicates Relative value measurement
2nd
display
Shows the reference value
1st
display
Shows the delta between the current measurement
data and the reference value
Manually set
the reference
value
1. To set the reference
value manually, press
the Shift key followed
by the REL key. The
setting appears.
REL
Indicates Relative measurement
1st display
Shows the reference value
2nd
display
Indicates Relative value modification