Jenway 6285 User Manual

Page 14

Advertising
background image

NAME

– user entered method name (up to a maximum of 19 characters)











EXCITATION NM

– user set excitation wavelengths. Actual filter data is checked

against this setting during measurement. If an incorrect filter is detected a warning
will be displayed.
EMISSION NM

– user set emission wavelengths. Actual filter data is checked against

this setting during measurement. If an incorrect filter is detected a warning will be
displayed.

Note:

If the excitation or emission filter modules do not match the current method

settings on entry to the measure mode (or prior to taking a reading or calibration)
then a warning is displayed:









A scrolling message will appear ‘please change method setting or fit correct filter
module

’. No measurements can be performed until this has been corrected. If a filter

is missing the instrument will not allow measurement to take place. Pressing the
EXIT key will return the screen to the setup menu options.
UNITS

– allows user to select the units appropriate to the test. ppm, RFU, U/ml,

mU/l, U/l, µM/l, mM/l, M/l, µg/ml, mg/ml, µg/dl, mg/dl, g/dl, ng/l, µg/l, mg/l, g/l, none,
%, ppb
RESOLUTION

– 0.01, 0.1, 1, 0.001

CALIBRATION SETUP

– this menu is not displayed when RFU (Raw Fluorescence)

units are selected.









EXIT

– allows the user to return to the previous menu

NO OF STDS

– allows selection of 1-6 standards (blank is always included)

STANDARDS TABLE

– user entered calibration standards

(11)