FLUORESCENCE TABLE

– shows results of last calibration RFU (raw fluorescence)

values. Tabular and graphical displays available












P.M.T GAIN(%)

- 0 to 100% gain setting for system photomultiplier tube. Note: This

option is not available on the Model 6270.
AUTO SET GAIN

- uses the highest user entered calibration standard to

automatically set the PMT gain to give as close as possible to 95% full scale
emission signal. Note: This option is not available on the Model 6270.
CAL REMINDER

- enables or disables the measure mode prompt to calibrate at the

set calibration interval
CAL INTERVAL(HRS)

- the number of hours between calibrations can be user set

from 1-999 hours.


Results Storage Menu

From the MAIN MENU press the down arrow key to move the highlight on the display
to cover RESULTS STORAGE, then press the enter key. The following display will
be shown:











EXIT

- allows the user to return to the previous menu

TIMED READ

- enables automatic measurements to be taken at a set interval when

in measure mod
INTERVAL

- time interval between the end of one reading and the automatic start of

the next reading in seconds
AUTO SAVE

- when enabled, automatically saves to instrument memory after each

reading is performed (either manually or timed)
AUTO PRINT

- when enabled, automatically prints to RS232 connected device after

each reading is performed (either manually or timed)