Jenway 6285 User Manual

Page 36

Advertising
background image

Section 15

Specification & Data


10.1 Technical Specification


Specification

6270 6280 6285

Light source

Pulsed Xenon Lamp with press-to-read operation

Expected Lifetime

No replacement required in normal operation

Sensitivity (Quinine Sulphate)

1µg/ml

<1ng/ml

<1ng/ml

Dynamic Range (Quinine
Sulphate)

5.2x 10

4

5.2 x 10

6

5.2 x 10

6

Detector Photodiode

Photomultiplier Photomultiplier

Wavelength Range

190 to 1100nm

190 to 650nm

190 to 850nm

Wavelength Selection

Intelligent Filter Modules (IFM)

Raw Fluorescence

a

a

a

Calibration

Blank + %

Blank + %

Blank + %

Concentration

a

a

a

Calibration

Blank + up to 6

standards

Blank + Up to 6

standards

Blank + up to 6

standards

Kinetics (Timed Readings)

In both above

modes

In both above

modes

In both above

modes

Time Limits

1 reading every 3 to 999 seconds

Methods

Stored

20 20 20

Results

Stored

100/method 100/method 100/method

GLP

Method lock with password access

Time and date stamped results

Last calibration date and time stored

Calibration Interval set

Calibration due reminder

User ID

Outputs

RS232 Serial Port

PC Software

Compatible with DataWay (option)

Sample Handling

10mm Square cuvette holder (standard)

Heated Sample Holder (Option)

Sipper Pump (Option)

Power Supply

90 to 230Vac universal mains adapter

12Vdc input (cigarette lighter cable and battery pack options)

Size

365(w) x 272 (d) x 160(h) mm

Weight 6Kgs












(33)