Settings 6.4.1. lan settings, Measurement parameter settings, Measurement parameter explanatory drawing – KANOMAX 3910 Portable Particle Counter Software Manual User Manual
Page 26

25
6. Measuring Software
6.4. Settings
6.4.1. LAN Settings
Click [Setting] -> [R LAN Setting…] to display the “LAN
Setting” dialog box.
Enter the Host Name of the instrument you wish to connect
with and click [OK].
You can also set the Host Name using [Search] (Refer to 3.5
Search Function).
6.4.2. Measurement Parameter Settings
Click [Setting] -> [Measurement Parameter Setting…] to
display “Measurement Parameter Setting” dialog box.
Configure the remote measurement parameter settings in the “Measurement
Parameter Setting” dialog box shown on the right. Refer to the settings table
below.
Setting Item
Setting Contents
Sampling Time
Set sampling time in seconds.
Key input: Integer 6~3600
Select: 6, 8, 10, 30, 60, 300, 600, 1200
Interval Time
Set interval time in minutes.
Key input: Integer 1~1440
Select: 1, 5, 10, 20, 30, 60, 120, 180
The Number of Samples
Key input: Integer 1~30000
Select: 1, 10, 60, 120, 300, 600, 1200, 3000
The Unit of Particle
Set the unit of particle to be used; p, p/cf or m
3
The Unit of Temperature Set the unit of temperature to be used;
deg C or deg F
The Unit of Velocity
Set the unit of air velocity to be used;
m/s or FPM
(Please refer to your instrument manual for further information on measurement settings.)
<Measurement Parameter Explanatory Drawing>
Start
Measurement
Finish
Measurement
Wait Time
Interval
Sampling Time
Interval
Sampling Time
The 1
st
Measurement
The 2
nd
Measurement
The 20
th
Measurement
Sampling Time
<Measurement Parameter Setting>