6. Measuring Software

6.4. Settings

6.4.1. LAN Settings

Click [Setting] -> [R LAN Setting…] to display the “LAN
Setting” dialog box.
Enter the Host Name of the instrument you wish to connect
with and click [OK].
You can also set the Host Name using [Search] (Refer to 3.5
Search Function
).


6.4.2. Measurement Parameter Settings

Click [Setting] -> [Measurement Parameter Setting…] to
display “Measurement Parameter Setting” dialog box.
Configure the remote measurement parameter settings in the “Measurement
Parameter Setting” dialog box shown on the right. Refer to the settings table
below.

Setting Item

Setting Contents

Sampling Time

Set sampling time in seconds.
Key input: Integer 6~3600
Select: 6, 8, 10, 30, 60, 300, 600, 1200

Interval Time

Set interval time in minutes.
Key input: Integer 1~1440
Select: 1, 5, 10, 20, 30, 60, 120, 180

The Number of Samples

Key input: Integer 1~30000
Select: 1, 10, 60, 120, 300, 600, 1200, 3000

The Unit of Particle

Set the unit of particle to be used; p, p/cf or m

3

The Unit of Temperature Set the unit of temperature to be used;

deg C or deg F

The Unit of Velocity

Set the unit of air velocity to be used;
m/s or FPM

(Please refer to your instrument manual for further information on measurement settings.)

<Measurement Parameter Explanatory Drawing>







Start
Measurement

Finish
Measurement

Wait Time

Interval

Sampling Time

Interval

Sampling Time

The 1

st

Measurement

The 2

nd

Measurement

The 20

th

Measurement

Sampling Time

<Measurement Parameter Setting>