background image

4

ELECTRICAL CONNECTIONS

44

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

4.7.4 Modular inputs/outputs and bus systems

Current output active (only current output terminals C/C- have HART

®

capability),

modular I/Os

• U

int, nom

= 24 VDC

• I ≤ 22 mA
• R

L

≤ 1 kΩ

• X designates the connection terminals A, B or C, depending on the version of the signal

converter.

Current output passive (only current output terminals C/C- have HART

®

capability),

modular I/Os

• U

ext

≤ 32 VDC

• I ≤ 22 mA
• U

0

≥ 1.8 V

• R

L, max

= (U

ext

- U

0

) / I

max

• X designates the connection terminals A, B or C, depending on the version of the signal

converter.

CAUTION!
Observe connection polarity.

INFORMATION!

•

For further information on electrical connection refer to Description of the inputs and outputs
on page 35

.

•

For the electrical connection of bus systems, please refer to the separate documentation for
the respective bus systems.

INFORMATION!

•

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

•

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions: Shield connected via the cable terminals in the terminal
compartment.

•

Any connection polarity.

Figure 4-12: Current output active I

a