Operation – KROHNE OPTISONIC 3400 EN User Manual

Page 86

Advertising
background image

6

OPERATION

86

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

C

C Configuration

Flow Simulation Active

Simulation of volume flow,

mass flow.

Switch off measured value

simulation.

VoS Simulation Active

Simulation of a certain

velocity of sound (VoS)

Switch off measured value

simulation.

Fieldbus Sim. Active

The simulation function in

the Foundation Fieldbus

module is active and being

used.

Check Fieldbus settings.

PROFIBUS Sim. Active

The simulation function in

the PROFIBUS module is

active and being used.

Check PROFIBUS settings.

IO A Simulation Active

IO A simulation is active.

Switch off simulation.

IO B Simulation Active

IO B simulation is active.

IO C Simulation Active

IO C simulation is active.

IO D Simulation Active

IO D simulation is active.

C

C Process

S

S Sensor

S

S Electronics

Electr.Temp.A Out of Spec

Temperature of the signal

converter electronics is out

of range.

Protect signal converter

from process effects and

sunlight.

Electr.Temp.C Out of Spec

Converter zero too large

Converter zero too large

Recalibrate the converter or

contact manufacturer

Advertising