6.3.4 Set free units
C5.8 HART
C5.8
HART
C5.8.1
HART
Switch HART
®
communication on/off:
Select:
On (HART
®
activated) possible current range for current output 4…20 mA /
Off (HART
®
not activated) possible current range for current output
0…20 mA
C5.8.2
Address
Set address for HART
®
operation:
Select: 00 (Point-to-Point operation, current output has normal function,
current = 4…20 mA) / 01...15 (Multi-Drop operation, current output has a
constant setting of 4 mA)
C5.8.3
Message
Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
C5.8.4
Description
Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
C5.8.5
HART long Tag
Up to 32 digits
C5.9 Quick Setup
C5.9
Quick Setup
Activate quick access in Quick Setup menu:
Select: Yes (switched on) / No (switched off)
C5.9.1
Reset Totaliser 1
Reset Totaliser 1 in Quick Setup menu?
Select: Yes (activated) / No (switched off)
C5.9.2
Reset Totaliser 2
Reset Totaliser 2 in Quick Setup menu?
Select: Yes (activated) / No (switched off)
C5.9.3
Reset Totaliser 3
Reset Totaliser 3 in Quick Setup menu?
Select: Yes (activated) / No (switched off)
Free units
Sequences to set texts and factors
Texts
Texts
Texts
Texts
Volume flow, mass flow and density:
3 digits before and after the slash
xxx/xxx (max. 6 characters plus a "/")
Permissible characters:
A…Z; a…z; 0…9; / - + , . *; @ $ % ~ () [] _
Conversion factors
Conversion factors
Conversion factors
Conversion factors
Desired unit
= [unit see above] * conversion factor
Conversion factor
Max. 9 digits
Shift decimal point:
↑ to the left and ↓ to the right