KROHNE ALTOSONIC V EN User Manual

Page 17

Advertising
background image

ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 17 of 106

4.1

Main menu: F1 Main window

The Main window is the default start-up window. This window shows an overview of the system and

can always be accessed by function key F1.

Explanation of the Main window layout:

UFC-DATA shows:
•

Raw data of the 5 channels regarding flow % and Velocity Of Sound (V.O.S.)

•

A red marker (

▪

) per channel shows an active channel failure, a green marker (

▪

) shows a

previously occurred channel failure

CONDITIONS show:
•

Temperatures, pressures and densities measured or calculated for the conditions of Proces,

Standard, Densitometer. Body temperature is also included.

•

A red marker (

X

) in front of a parameter shows an alarm for out of range or manual on-line

override, a green marker (

X

) shows a previously occurred alarm

UFP-CALC shows:
•

Flow rates at Proces conditions, Standard conditions and Mass

.

RESETABLE TOTALISERS shows:
•

The forward, reverse and summation of the Totaliser values at Proces conditions, Standard

conditions and Mass.

•

The resetable totalisers can be reset in the Control menu (through F9 in main): F8 RES-TOT.

It is also possible to reset the totalisers by digital input signal or Modbus boolean.

NON RESETABLE TOTALISERS shows:
•

The forward, reverse and summation of the Totaliser values at Proces conditions, Standard

conditions and Mass.