REAL PROFILE:

On GENERAL FLOW error “1-4 channels down” the REAL profile is used to correct the channels with

errors. The real profile is sampled at a certain flow rate.
•

The REAL-profile correction has a limited validity.

When the actual profile changes too much, the previously sampled REAL-profile might not be

reliable anymore. The check for profile changes is done through flow-rate difference.

•

When the sampled REAL-profile flow rate differs too much from the actual flow-rate during REAL-

profile correction this is shown as a warning.

CORRECTION WARNINGS shows:
•

If there is too much flow variation for corrections, the corrections go on hold. When the corrections

are on hold the real time profile is used as a standard for correcting the flow.

•

If there are too much flow variations or channels failing, the sampling of the REAL profile goes on

hold. On release the sampling is started at maximum time for sampling a profile.

SYSTEM ERRORS shows:

The status of the system is divided into:
•

System Runtime Warnings. These are caused by system failures. These failures will not influence

the flow measurement.

•

System Runtime Alarms. These are caused by system failures. These failures might influence the

flow measurement.

Identified System Runtime Errors are numbered 1 to 60 are:

4.2.1

SYSTEM ERRORS

Identified System Runtime Errors are numbered 1 to 60, A = alarm, W = warning:

Error no. In function

Problem

Consequence

A : 1

Get RS485 data from

converters

Overrun, missed data

Missed data, message

A : 2

Self test

Error in memory self-test

Non-reliable memory

A : 3

Batch start / stop

Error during saving files of start or stop

File lost but ticket is made

A : 4

Profile correction (REAL)

Error in state_correction

Attempt divide to by zero

W: 5

Read Backup all files

Error in reading backup file

Possible loss of backup file

W: 6

Switching disk

Error in finding a drive

Message

W: 7

System time

A notice that the system time was adjusted
manually or by Modbus.

No consequence for totalisers or
proces time, only on ticket time

A: 8

General program

exectutable certification

CRC of the executable is not correct

executable file is corrupted.

Load a new executable file.

Contact KROHNE service for

help

A: 9

Batch status backup

Status file corrupt

Possible loss of batch status

W: 10

Override values files

Error in opening/closing override value file

Override values not stored but

still in use

A: 11

Batch totaliser backup

Totaliser backup-file corrupt

File lost , message

A: 12

Batch average backup

Average backup-file corrupt

File lost, message

A: 13

Batch ticket create

Error in creating batch ticket file

Ticket itself is made for printing

but lost during saving

W: 14

Opening file (for update)

Error in opening REAL file

File lost, message

W: 15

Closing file (for update)

Error in closing REAL file

File lost, message

W: 16

API settings

Error in file, defaults are loaded and saved

Old settings lost

W: 17

Batch 2

A alarm on batch 2 file (Batch 2 is only
used through Modbus with a Scada

system)

File lost, message

W: 18

Check free disk-space

Error dos_getdiskfree() call

Time-out function 30 s

W: 19

Check free disk-space

Low on disk-space

Time-out function 30 s

W: 20

Ad card overrun

The requested AD card is not noticed

Solve the problem

W: 21

Opening file (for update)

Error opening API table file

File lost, message

W: 22

Value check

1 or more API values defaulted

Check the installed parameters

W: 23

Opening file (for update)

Error opening external flow meter file

File lost, message

W: 24

Value check

Default external flow meter K-factor

Check the installed K-factor

W: 25

Counter input

Unable to read Counter value

Read on next entry

A : 26

Calibration MP103 card

MPCA File corrupt

Install backup