ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 81 of 106

8 OUTPUT

The output consists of:


•

Frequency output MP103 card

•

Analog output MP103 card

•

Relay outputs MP103 card

•

Analog outputs AD card

•

Digital outputs AD card

•

Modbus communication

8.1

Frequency output MP103 card

Frequency output:

-

Maximum output range is software adjustable 1 – 2000 Hz

-

12V/24V / open-collector selectable by card jumpers

-

There is one output value but there are two physical outputs, these can be phase-shifted 90°/180°

selectable by card jumper to simulate a turbine output for pulse fidelity and integrity check.

The resolution of the frequency output is max 0.016% of the output value. The resolution mentioned is

for a static output value. In practice, the output resolution will be averaged because of the variations in

signal. Over a period of time with different output values the resolution will not be an issue.

The most likely frequency output is the proces volumetric flow (default).
•

The frequency output function can be disabled/enabled in the Initialisation file: HSET0300.ufp

section 3

•

The frequency output can be configured in the Initialisation file: CLNT0300.dat section 5

•

The signals can be checked on value in the service window: IO.

•

Monitoring is also possible by its calibration program (see Manual: ALTOSONIC V UFP

Calibration and Verification I/O)

HSET0300.ufp section 3

CLNT0300.dat section 5

-----------------------------------------------------------------------------
5 <FREQUENCY OUTPUT, mp103 card>
5.1 Freq_max =#1000 //Max.scale [Hz], range= 1 - 2000 [Hz]
5.2 Freq_mode =#1 //0=DIS 1=flow[m3/h] 2=flow15 3=mass[ton/hr]
//4=dens[kg/m3] 5=c_s[m/s] 6=VCF 7=viscosity[10e-6
m2/s]
//8=dens15[kg/m3] 9=Temp[øC] 10=Pres[bar]
5.3 Freq_min_unit =#0 //Min outputvalue in [unity]
5.4 Freq_max_unit =#1800 //Max outputvalue in [unity]
5.5 Freq_tau =#0 //Averaging time tau[s]
5.6 Freq_dir_flow =#1 //Directionflow for output frequency: 0=+, 1=+-

3.1 MP_freq_out =#0 //Frequency output 0=disable, 1=enable