KROHNE ALTOSONIC V EN User Manual
Page 70

ALTOSONIC V
Operating Manual rev02 E
7.30850.34.00
Page 70 of 106
•
Now batch is ready to start by Function key F10 or by Modbus command if enabled. Note that
depending on security level it is only possible to start a batch if flow is at zero flow conditions
•
Possible to cancel the set-up (F2)
•
Or return to the API settings (F3) or the Text settings (F4)
6.9
Batch start
Starting a batch holds the following automatic actions:
•
Reset of: errors, resetable totalisers and batch flow-weighted averages (temp, press, densities
etc.)
•
Increase ticket number by one (is saved in the “batch status” file.
•
Saving of all batch parameters (for later use when batch is stopped and certain batch start values
are requested on ticket) in a “batch start” file that is secured by a CRC-checksum
New since software version 03.00.50.01 is the option to view, during the batch, the worst case batch
Volume Error % estimate due to batch alarms such as path failures, input signals alarms etc etc