Now batch is ready to start by Function key F10 or by Modbus command if enabled. Note that

depending on security level it is only possible to start a batch if flow is at zero flow conditions

•

Possible to cancel the set-up (F2)

•

Or return to the API settings (F3) or the Text settings (F4)

6.9

Batch start

Starting a batch holds the following automatic actions:
•

Reset of: errors, resetable totalisers and batch flow-weighted averages (temp, press, densities

etc.)

•

Increase ticket number by one (is saved in the “batch status” file.

•

Saving of all batch parameters (for later use when batch is stopped and certain batch start values

are requested on ticket) in a “batch start” file that is secured by a CRC-checksum

New since software version 03.00.50.01 is the option to view, during the batch, the worst case batch

Volume Error % estimate due to batch alarms such as path failures, input signals alarms etc etc