ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 66 of 106

3

At all flowing conditions

No

Yes (continuous pipe

line measurement)

4

At all flowing conditions

No

Yes (continuous pipe

line measurement)

BATCHING_ON 1 and 2 have the following restrictions during a batch:

-No reset possible of resetable totalisers

-No reset of error times but the possibility to reset occurred error messages


•

The previous number of tickets saved is set with MAX_TICKETS. Default is 100 tickets.

Be careful with increasing the number of tickets. Not enough disk space means losing tickets


•

For Continuous Pipe Line Measurement the ticket automatically is printed starting from

HOUR_START

•

For Continuous Pipe Line Measurement the ticket automatically is printed every

HOUR_INTERVAL, but if interval 0 is installed than tickets are only printed on demand

•

By MOD_BUS_CONTROL it is possible to enable the controls through Modbus for batching:

-Start batch

-Stop batch

-Reset printing

-Confirm printing

Or in case of using continuous Pipe Line measurement

-Ticket on demand with reset of values

-Ticket on demand without reset of values

- Reset printing

6.5

Batch status


Batch status
(status window text)

As a value on
Modbus

Explanation

NON

0

No batch active, ready to set-up

SETUP

1

In set-up mode. After set-up is done, it is possible to

start a batch

RUNNING

2

Batch is started

END-BATCH

3

Batch is stopped and ticket is made, then attempt to

END_PRINT

END-PRINT

5

Status during successful printing

END-FAIL

6

If printing fails or printer is busy too long

CONFIRM

7

After successful print job waiting for manual

confirmation

RESET

10

Waiting for reset command after END_FAIL