ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 24 of 106

that the factor is used in the flow calculation. The correction is lineair and depends on meter

construction.

For example correction per 100 Bar difference for:

•

a certain ALTOSONIC V 6 inch construction is about +0.04%

•

a certain ALTOSONIC V 10 inch construction is about +0.045%

As this is a linear function the correction at 50 Bar difference would be half the value (0.02%).

This was calculated using the standard: ISO/CD 17089/1

The correction is described in different standards. To be compliant with these different standards

KROHNE as made the choice to put the correction in a general formula:

)

(

100

1

ref

proces

pb

pb

P

P

C

K

−

⋅

+

=

K

bp

: Correction factor used for the pressure expansion

C

bp

: Lineair pressure correction factor [%/bar], depending on the construction and used standard.

P

proces

: Pressure proces[bar]

P

ref

: Pressure reference [bar],

Note that P expansion correction is disabled at that time, or used pressure is low.

STANDARD VOLUME CORRECTIONS shows:
•

The temperatures, pressures and densities at proces, standard, densitometer and optional

external flow meter conditions in relation to the correction factors Ctl and Cpl

•

The correction factors Ctl (temperature correction to 15°C) and Cpl (pressure correction to

1.01325 bar, or 0 barg)

See chapter 5 for more information on the Standard Volume correction