SERVICE
7
123
MFC 300
www.krohne.com
02/2012 - 4000498602 - MA MFC 300 R03 en
7.1.2 Replacing the signal converter electronics (BE)
• Remove the front panel. Use a small screwdriver to open the plastic clips that hold the display.
• Remove the 2 locking screws.
• Remove the electronics by pulling sideways on the plastic housing.
i
The signal converter electronics slide out easily once they have become detached from the
backplane PCB.
• To insert the new electronics, slide them back into the housing, retighten the two screws and
attach the display.
The measuring system recognises the hardware replacement when the power supply is
switched on. Following replacement of the sensor electronics (SE), or the entire sensor
including electronics or the signal converter electronics (BE), the device indicates a "fatal" error
status. In this case, the menu features various options depending on the status detected.
DANGER!
Work on the signal converter electronics may only be performed when disconnected from the
power supply.
.book Page 123 Monday, February 6, 2012 11:01 AM